Meriam's M2 Series Smart Manometers bring high precision and value to handheld, digital manometer users. The M2 features NIST traceable accuracy of ±0.025% of F.S., independent of temperature effect.  Pressure ranges from 10" H2O F.S. to 3000 PSIG F.S. are available.

M2 pressure sensors are available to measure gauge, differential, absolute or vacuum pressure. The display can read out in any of eleven pressure units or can be linearly scaled for special user units. Differential models offer a square root flow function to display flow rate in user defined units. An adjustable damping feature minimizes the effects of pulsating pressures. A Min/Max function captures the extremes of pulsating or varying pressure signals and a Hold function can freeze the display at any point of interest. Auto Record documents up to 240 readings for future recall.

Media Compatibility

DN: Non-isolated differential sensor for clean, dry, non-corrosive gases. GI & AI: Isolated Gauge or Absolute sensor for fluids compatible with 316L SS. See Meriam's M200-DI for wet/wet DP requirements.

Pressure Limits

Twice range on Gauge and Absolute (GI and AI) units. Twice range on Differential (DN) units when pressurized on high side only and 150 PSI (10.5 Kg/cm2) static when applied to both sides of the sensor simultaneously.

Accuracy:

+/- 0.05% F.S. or optional +/- 0.025% F.S.

Engineering Units:

Inches of H2O (select 4ºC, 20ºC, or 60ºF), PSI, mm of Hg, inches of Hg, mBars, Bars, kPa, kg/cm2, cm of H2O and user selectable scaling.

Pressure Differential Gauge Absolute
Units "H2O PSI mmHg
Range 0-10 0-2 0-17.403
Range 0-28 0-15 0-38.674
Range 0-200 0-20 0-100
Range 0-415 0-30 0-1000
Range 0-2000 0-50 0-2000
Range - 0-100 -
Range - 0-200 -
Range - 0-300 -
Range - 0-500 -
Range - 0-1000 -
Range - 0-3000 -
Pressure Limits 2 x Range 2 x Range 2 x Range

Certifications

CE compliant. Optional Intrinsically Safe (I.S.) per ATEX and cULus

We periodically send an email newsletter containing updates, new product releases, training and advice. If you would like to receive these please subscribe below.

Join

Leak Test

Allows users to view Min/Max pressure values and calculates leak rate per minute in pressure units. The feature is very handy when troubleshooting or pressure testing pneumatic systems.

Auto Record

Document up to 240 readings (automatically or manually) to reduce paperwork and field time. Saved readings can be recalled at any time.

Damping Rates

Exponential damping with user selectable time constant from 0.1 to 25 seconds.

Min/Max Capture

Capture speed is equal to the selected damping rate.

Field Re-cal

The M2 series can be recalibrated in the field for zero, span, and linearity. The user has the option to select a 1 point or 5 point calibration procedure. Enables users with accurate primary standards to calibrate their devices locally if desired.

Auto Shut-Off

User selectable shut-off. Programmable at 10, 20, 30, 45, 60 minutes.